3.5.
CABINET EXHAUST DUCTING
Ducting of the exhaust may be preferred to exhaust in the lab in order to, for example:
-
Prevent from lab temperature rise.
-
When volatiles and toxic vapours are involved during the process.
WARNING
Ducting the exhaust to the outside of the lab creates a depression in the lab.
3.5.1.
OUTSIDE DUCTING VIA A CANOPY
In case of direct ducting outside it must be installed and additional fan to the duct, a canopy,
and adjust the velocity parameters. GS has not been tested. It must be tested after
installation.
WARNING
When the cabinet exhaust is connected to an assisted/air forced exhaust system, it is
necessary to install a canopy between the cabinet and the exhaust duct. This will prevent
from the loose of operator protection in case of any decrease of the exhaust airflow system
(duct obstruction, blower control failure…).
See description & installation in
Annex A4.4.
3.6.
INSTALLATION TESTS
According to standard EN 12469 and in order to ensure correct cabinet operation, after the
installation and connection of the cabinet, it is compulsory to make the following tests
(before the equipment is used for the first time)
-
9.1. Impulsion filters integrity Test.
-
9.2. Exhaust filters integrity Test.
-
9.3. Downflow air velocity.
-
9.3.1 Downflow set points alarm test.
-
9.4. Inflow air velocity Test.
-
9.4.1 Inflow set points alarm test.
-
9.5. Smoke test.
TELSTAR also recommends to do the following tests:
-
9.6. Light Test.
-
9.7. Noise Level Test
Telstar, its representatives and distributors put a qualified technical team at your disposition
that uses control equipment with maximum health protection for carrying out this installation
test under full safety guarantees.
Содержание Bio II Advance
Страница 2: ...MANUAL DE USUARIO Cabinas de Seguridad Microbiol gica Bio II Advance...
Страница 3: ...UM BIOIIADVANCE SP 1302 2...
Страница 35: ...UM BIOIIADVANCE SP 1302 34 11 CERTIFICADOS 11 1 CE Ver Anexo A10 1 11 2 GARANT A Ver Anexo A10 2...
Страница 36: ...USER S MANUAL Microbiological Safety Cabinet Bio II Advance...
Страница 37: ...UM BIOIIADVANCE EN 1302 2...
Страница 69: ...UM BIOIIADVANCE EN 1302 34 11 CERTIFICATES 11 1 CE See Annex A10 1 11 2 GUARANTEE See Annex A10 2...
Страница 70: ...MANUEL D UTILISATION Cabines de s curit microbiologique Bio II Advance...
Страница 71: ...UM BIOIIADVANCE FR 1302 2...
Страница 103: ...UM BIOIIADVANCE FR 1302 34 11 CERTIFICATS 11 1 CE Voir Annexe A10 1 11 2 GARANTIE Voir Annexe A10 2...
Страница 104: ...BEDIENUNGSANLEITUNG Mikrobiologische Sicherheitswerkb nke Bio II Advance...
Страница 105: ...UM BIOIIADVANCE DE 1302 2...
Страница 120: ...UM BIOIIADVANCE DE 1302 17 Abb 1 X Abb 2 X Abb 3 X Abb 4 X...
Страница 139: ...UM BIOIIADVANCE DE 1302 36 11 ZERTIFIKATE 11 1 CE Siehe Anhang A10 1 11 2 GARANTIE Siehe Anhang A10 2...
Страница 140: ...ANNEXOS ANNEXES ANNEXES ANH NGE Bio II Advance...
Страница 141: ...A1 BIOIIADVANCE SP EN 1302 1 Vlam Vent Qext SP Anexo A1 ESQUEMA DE PRINCIPIO EN Annex A1 PRINCIPLE SCHEME...
Страница 142: ...A1 BIOIIADVANCE FR DE 1302 1 Vlam Vent Qext FR Annexe A1 SCH MA DE PRINCIPE DE Anhang A1 FUNKTIONSPRINZIP...
Страница 156: ...A5 1 BIOIIADVANCE FR DE 1302 3 Fig 3...
Страница 183: ...A7 BIOIIADVANCE SP EN FR DE 1302 2...
Страница 184: ...A7 BIOIIADVANCE SP EN FR DE 1302 3...
Страница 185: ...A7 BIOIIADVANCE SP EN FR DE 1302 4...
Страница 186: ...A7 BIOIIADVANCE SP EN FR DE 1302 5...
Страница 187: ...A7 BIOIIADVANCE SP EN FR DE 1302 6...
Страница 188: ...A7 BIOIIADVANCE SP EN FR DE 1302 7...
Страница 189: ...A7 BIOIIADVANCE SP EN FR DE 1302 8...
Страница 190: ...A7 BIOIIADVANCE SP EN FR DE 1302 9...
Страница 191: ...A7 BIOIIADVANCE SP EN FR DE 1302 10...
Страница 226: ......
Страница 227: ......