H
Internal clock frequency
H
External clock input
H
Sequence & digital output
H
Inhibit function
H
Output level, amplitude, rise time and fall time accuracies
H
Delay time accuracy
This test confirms the frequency accuracy of the DG2030 internal clock.
With the PLL circuit on: Clock frequency
±
0.0001 %
With the PLL circuit off: Clock frequency
±
3 %
Equipment Required
Frequency counter
BNC cable
50
W
termination
NĆtoĆBNC adapter