Setup Menu
3Ć76
DG2030 User Manual
BK0 BK1 BK1 BK1 BK2 BK2 BK0 BK1 BK1 BK1 BK2 BK2
BK0
1
BK1
3
BK2
2
BK0
BK1
BK2
0
1
2
0
1
2
2
1
0
Line No.
Block
Repeat Count
Pattern Data
Line No.
Sequence
Output
Sequence
Figure 3Ć57: Repeat Mode pattern data output (when a sequence is defined)
Pattern data is output only once when a trigger input is received. A trigger signal
is received either when the front panel FORCE TRIGGER button is pressed or
when an external trigger signal is input to the TRIGGER INPUT connector.
When no sequence is defined, all of the pattern data is output as a single
waveform as shown in Figure 3-58. If a sequence is defined, the sequence
ordering and repeat (Repeat Count) functions are applied as shown in Figure
3-59. The extended sequence functions (trigger input, event jump, etc.) are
ignored.
Pattern Data Output
Trigger Input
Pattern Data
Figure 3Ć58: Single Mode pattern data output (when no sequence is defined)
Single