Appendix B: Performance Verification
BĆ20
DG2030 User Manual
The following steps check that the DG2030 internal trigger source generates a
1
m
s trigger event every 10 seconds.
25. Press the following buttons to set the DG2030 trigger interval to 1
m
s.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Interval
State (Set to On)
Time
1
MHz/
m
s
26. Change the frequency settings in the function generator to 1 MHz.
27. Change the oscilloscope horizontal setting to 200 ns/div.
28. Confirm that trigger event is generated in every 1
m
s with the oscilloscope
screen.
This can be made by confirming that output signal appears on the oscillo-
scope CH2 in every 1
m
s.
29. Press the following buttons to set the DG2030 trigger interval to 10 s.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Interval
State (Set to On)
Time
10
Hz/s/V
30. Stop acquisition in the oscilloscope, and then change the oscilloscope
settings to:
Horizontal axis
5 s/div
. . .
Record length
500
. . . .
Acquisition mode
Single Acquisition Sequence
.
(Stop After menu)
Trigger source
CH3
. . . .