Appendix B: Performance Verification
DG2030 User Manual
BĆ13
Setup.
H
Oscilloscope
Displayed channel
CH1, CH2, CH3 and CH4
Vertical axis
1 V/div for CH1 and CH2
. . . . .
5 V/div for CH3 and CH4
Horizontal axis
400 ns/div
. . .
Record length
5000
. . . .
Acquire mode
Peak Detect
. . . .
Single acquisition sequence (Stop After menu)
Trigger mode
Auto
. . . .
Trigger level
2 V
. . . . .
Trigger source
CH4
. . . .
Trigger position
3 %
. .
Input coupling
DC
. . . .
Input impedance
50
W
for CH1 and CH2
. .
1 M
W
for CH3 and CH4
H
Function generator
Waveform
Square wave
. . . . . . .
Frequency
500 kHz
. . . . . . .
Amplitude
1 V (50
W
termination))
. . . . . . .
Offset
500 mV
. . . . . . . . . .
Characteristics Confirmation Procedure.
1. Load the TP3PG.PDA test pattern file from the performance check disk.
The following steps check the pattern data output operation in Repeat run mode.
2. Press the following buttons to set the DG2030 run mode to Repeat.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Run Mode
Repeat
3. Run the oscilloscope for acquisition.
4. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
5. Confirm that the data pattern shown in Figure B-6 appears on the oscillo-
scope.
6. Press the START/STOP button on the front panel to stop output (the
button’s LED goes off).