Appendix A: Specifications
AĆ6
DG2030 User Manual
Table AĆ2: Warranted environmental characteristics
Characteristics
Description
Temperature
Operating
+10
_
C to +40
_
C
NonĆoperating
-20
_
C to +60
_
C
Relative humidity
Operating
20% to 80% (No condensation)
Maximum wetĆbulb temperature 29.4
_
C
NonĆoperating
5% to 90% (No condensation)
Maximum wetĆbulb temperature 40.0
_
C
Altitude
Operating
To 4.5 km (15000 ft).
Maximum operating temperature decreases 1
_
C each 300 m above 1.5 km.
NonĆoperating
To 15 km (50000 ft).
Dynamics
Vibration
Operating
0.27 Grms, 5 to 500 Hz
NonĆoperating
2.28 Grms, 5 to 500 Hz
Shock
Non operating
294 m/s
2
(30 G), halfĆsine, 11 ms duration.
Three shocksper axisin each direction (18 shockstotal)
Installation requirements
Power consumption
300 wattsmax. Maximum line current is4 A rmsat 50 Hz
Dissipation (fully loaded)
90 V libe, with 5 % clipping
Surge current
x
9 A peak for less than 5 line cycles, at 25
_
C, after product has been off for at least 30 s.
Cooling clearance
Top clearance
2.5 cm (1 inch)
Side clearance
15 cm
(6 inches)
Rear clearance
7.5 cm (3 inches)