Setup Menu
DG2030 User Manual
3Ć73
Sets whether all outputs should be set to the high-impedance state or whether
they should continue to output their current values when the output is stopped
with the START/STOP button. When Z on Stop is set to On, the output is set
to the high-impedance state.
Operation.
Set the output impedance when output is stopped
Bottom button
Popup menu
Side button
Level Condition
Z on Stop
(select On or Off)
Timing Condition Menu
This menu lets you set each channel’s rise, fall, and delay timing parameters.
Figure 3-55 shows the current rise and fall time settings of an example output
channel. You can also see the same image in the SETUP menu. Highlighted lines
indicate the pulse edges currently set. The three non-highlighted horizontal lines
from the top indicate the highest, 0 and lowest voltage levels that can be set. The
two non-highlighted vertical lines indicate the signal rising edge at left and
falling edge at right.
If the pulse edges are positioned at the non-highlighted vertical lines, the positive
delay is currently set to zero. If the pulse edges are positioned left of the
non-highlighted lines, the output pattern data will be delayed.
You can adjust the rise and fall times within this range indicated by two
non-highlighted slope lines or set to FAST. When you try to set a rise or fall time
out of the valid range, the display flashes the word Rise or Fall beneath the
menu pulse image.
The range varies depending on the pulse amplitude. See Figure A-1 on page A-4
for the relationship between range and pulse amplitude and the value represented
by FAST.