Appendix B: Performance Verification
BĆ34
DG2030 User Manual
DG2030 Front Panel
Clock Output
BNC Cables
TDS784A Front Panel
CH0
CH1 CH2
Figure BĆ13: Delay time measurement connections
Setup.
H
Oscilloscope
Displayed channel
CH1 and CH2
Vertical axis
500 mV/div
. . . . .
Horizontal axis
5 ns/div
. . .
Trigger mode
Auto
. . . .
Trigger source
CH1
. . . .
Trigger level
750 mV
. . . . .
Input coupling
DC
. . . .
Input impedance
50
W
. .
Characteristics Confirmation Procedure.
1. Load the TP8PULSE.PDA test pattern file from the performance check
disk.
2. Press the START/STOP button on the front panel so that the button LED
indicator turns on.
3. Press the following buttons to set the DG2030 CH0 output delay time to
–5.00 n.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Timing
Condition
Move the cursor to CH0 with the general purpose knob.
Delay
-5.0
ENTER