Appendix B: Performance Verification
BĆ14
DG2030 User Manual
CLOCK OUT
CH0
EVENT OUT
SYNC OUT
CLOCK OUT
CH0
EVENT OUT
SYNC OUT
BLOCK 2
BLOCK 4
BLOCK 1
BLOCK 3
BLOCK 1
BLOCK 3
BLOCK 2
BLOCK 4
Figure BĆ6: Repeat mode timing chart
The following steps check the pattern data output operation in Single run mode.
7. Press the following buttons to set the DG2030 run mode to Single.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Run Mode
Single
8. Run the oscilloscope so that the oscilloscope is waiting for a trigger event.
9. Press the START/STOP button on the front panel so that the button’s LED
indicator lights.
10. Confirm that the CH0 output data pattern as shown in Figure B-7 appears on
the oscilloscope screen once every time the FORCE TRIGGER button on
the DG2030 front panel is pressed.
11. Press the START/STOP button on the front panel to stop output (the
button’s LED goes off).