Edit Menu
DG2030 User Manual
3Ć17
When CSV format is selected, the instrument displays a menu for selecting
which bits should be written to the file, delimiter symbols, and other parameters.
Table 3-4 describes the parameter items that are set using this menu.
Table 3Ć4: Export parameters
Parameter
Function
Export Data Bit
Specifiesthe data bitsto be written to the pattern data. Bitsfor which
thisindicator ison are written. Bitscan be selected using the general
purpose knob, and the on/off state can be changed with the left and right
arrow buttons. Consecutive bits can be turned on or off using the 1 and
0 numeric keys.
Column separator
Sets the line separator symbol. The delimiter symbols are usually used
asfollows:
CR+LF: MSĆDOS and Windows
LF: UNIX
CR: Macintosh
Data separator
Sets the interĆbit delimiter. Comma, space, or tab can be selected for
this parameter. Comma is the most common setting.
From
Thisfield isvalid when Entered isspecified for the Region in the sub
menu. It specifies the starting position for the data written.
Size
Thisfield isvalid when Entered isspecified for the Region in the sub
menu. It specifiesthe number of data pointswritten.
When the AWG2000 Series format is selected, a menu that allocates data for a
total of 14 bits, MARKER 1 and 2 and DATA00 to DATA11, is displayed. Use
the arrow buttons to select the AWG2000 Series bit and use the general purpose
knob to allocate the DG2030 bit. Allocate all bits to be written by repeating this
operation.
Sub Menu
Item
Function
Set All Data bits
(CSV only)
Setsthe data bit write settingsfor all data bits.
Clear All Data bits
(CSV only)
Clearsthe data bit write settingsfor all data bits.
Region
When thissetting isset to All, the whole data area iswritten, and when
it is set to Entered, the data in the area specified by From and Size is
written.