Operating Examples
2Ć48
DG2030 User Manual
This step actually outputs the signals and confirms those signals with an
oscilloscope.
9. Connect the outputs to the oscilloscope.
Connect the CH0 and CH1 on the front panel to the oscilloscope channels
CH1 and CH2. This requires two BNC to BNC cables as shown in Figure
2-29.
DG2030
Oscilloscope
Two BNC Cables
Figure 2Ć29: DG2030 and oscilloscope connection
10. Press the START/STOP button on the front panel.
11. Set up the oscilloscope appropriately and observe the pattern signals on the
oscilloscope screen.
12. Save the state of the settings on the floppy disk.
a. Insert an IBM-formatted floppy disk in the drive.
b. Name the new file COUNT3.
Menu button
Bottom button
Popup menu
Side button
Front panel
button
EDIT
File
Save Data &
Setup
Clear String
COUNT3
OK
The data in this procedures will be saved in the file COUNT3.PDA.
Signal Output
Saving State of the
Settings