Appendix B: Performance Verification
BĆ10
DG2030 User Manual
This test confirms external clock input operation. Confirm that the clock signal is
output from the CLOCK OUT connector when a square wave signal with a
frequency of 1 MHz and an amplitude of 1 V (V
IH
> 0.7 V, V
IL
< 0.3 V) is input
as an external clock signal.
Equipment Required
Oscilloscope
Function generator
Two BNC cables
Performance check disk
Connections.
Connect the clock output from the DG2030 front panel to the
oscilloscope CH1 input with an BNC cable. Connect the function generator
output to the DG2030 rear panel clock input with the other BNC cable.
DG2030 Front Panel
Clock Output
BNC Cable
TDS784A Front Panel
CH1 Input
AFG2020 Front Panel
BNC Cable: connect to the
AUX INPUT:
CLOCK
connector on the DG2030 rear panel.
Figure BĆ4: External clock input connections
External Clock Input
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com