Appendix B: Performance Verification
BĆ6
DG2030 User Manual
Table BĆ3: Error codes
Error code
Error meaning
Failed component
1XX
CPU diagnostics error
A6 CPU board
2XX
Display diagnostics error
A6 CPU board
3XX
Front panel diagnostics error
A12 Key board
4XX
Clock diagnostics / calibration error
A30 Clock board
5XX
Trigger diagnostics / calibration error
A30 Clock board
6XX
Sequence memory error
A50 PG&Output board
7XX
Pattern memory diagnostics error
A50 PG&Output board