Appendix B: Performance Verification
BĆ18
DG2030 User Manual
H
The event signal pattern in the AUX OUTPUT: EVENT outputs
correctly, and the event high and low levels are nearly 5 V and 0 V,
respectively.
H
The sync signal pattern in the AUX OUTPUT: SYNC outputs correctly,
and the sync high and low levels are nearly 5 V and 0 V, respectively.
H
Change the oscilloscope horizontal setting to 2 ms/div, and confirm that
the trigger wait function operates correctly.
Trigger event is generated in every 10 ms. Confirm that the block pattern
(BLOCK 1) waits for the trigger event and then outputs.
22. Press the following buttons to set the DG2030 clock frequency to
409.6 MHz.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Clock
Int FREQ
409.6
MHz/
m
s
23. Change the oscilloscope horizontal axis setting to 10 ns/div.
24. Confirm that the data pattern in the CH0 to CH3 (also CH4 to CH7 if the
option 01 has been installed), EVENT OUT and SYNC OUT can be
observed as shown in Figure B-8 on the oscilloscope. Change the channel
output connection as needed.