SVAN 979 User Manual
Gradient trigger
The
Grad +
trigger starts the 1-second measurement/integration on condition:
value of the measured result (
Source
) averaged by 0.5 ms is greater than the
threshold value (
Level
) and the gradient of this Source is greater than the
gradient threshold value (
Gradient
). In other cases, the instrument continues
checking the trigger condition every 0.5 ms.
This trigger has the same logic as the
Level +
trigger, but the trigger condition requires also the gradient
threshold level to be exceeded.
External trigger
When
External
is selected, the measurement is triggering by the external signal
on the
I/O
socket. In this case, it is necessary to set up the
I/O Mode
parameter
as
Digital In
(
path:
<
Menu> / Instrument / Multifunction I/O
).
After the measurement/integration start from the external trigger, the
measurement/integration will continue for the
Integration Period
.
Note: When a measurement is waiting for the gradient or external
trigger the “trigger” icon appears alternatively with the „measurement”
icon.
Source result
Only one measured result (
Source
) can be used for checking trigger condition in the
Level Meter
mode, namely
the instantaneous RMS from the first profile (with appropriate filter and detector), which is denoted here as
Leq(1)
in the sound mode and
RMS(1)
in the vibration mode. This position cannot be changed.
Threshold level
The threshold level for the measurement
trigger (
Level
) can be set with 1 dB step in
the range from 24 dB to 136 dB in sound
modes and from 64dB (1.585mm/s
2
) to
176dB (631 m/s
2
) in vibration modes.
Rate of Source value changing
This position appears when the
Grad+
trigger is chosen. The rate of the Source
value changing (
Gradient
) can be set in the range from 1 dB/ms to 100 dB/ms.
4.3 S
ETTING PARAMETERS FOR PROFILES
–
P
ROFILES
The
Profiles
list enables programming of
three virtual level meters (Profiles), which
measure the set of results with the same
weighting filter (
Filter
) and exponential
detector time constant (
Detector
) (see
Appendix D). Default values
– see
<ENT>
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...