![Svantek SVAN 979 Скачать руководство пользователя страница 117](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285117.webp)
117
SVAN 979 User Manual
12
REVERBERATION TIME MEASUREMENTS
– RT60
The
RT60
analysis is an optional function of SVAN 979, which provides reverberation time calculation for
1/1-octave bands (from 63 Hz to 16 kHz) or 1/3-octave bands (from 50 Hz to 20 kHz) and three total RMS
levels (
A
,
C
and
Z
weighted). Whole measurement process and calculations implemented in SVAN 979 fulfil
the ISO 3382 standard.
The reverberation time of the room can be obtained with the use of SVAN 979 by two measurement methods:
Impulse Response Method
(
Impulse
) and Interrupted Noise Method
(
Decay
). The selection of the method
depends on the type of the used sound source. The
Impulse
method is designed for measurements using the
impulse sound source (like pistol shot, petard explosion), whereas the
Decay
method is intended for
measurements when room is excited by the broad or narrow band sound noise source (usually pink noise). For
more details about the measurement and calculation process see Appendix E.
The reverberation time analysis applied in the instrument consists of two parts:
1. The measurement part during which the acoustic response of the room is registered.
2. The calculation part during which the reverberation time (
EDT
,
RT20
,
RT30
and
RTUser
) is calculated
for the measured room response.
Note: It is recommended to familiarize with Appendix E before proceeding. This chapter describes
only the navigation of the instruments, whereas Appendix E depicts the definitions and describes
reverberation time measurements.
12.1 S
ELECTING THE
RT 60
FUNCTION
To activate the
RT60
analysis function,
select
the
RT60
position
in
the
Measurement Function
list and press
<ENTER>
.
<ENT>
Note: The RT60 function is optional and should be unlocked by entering the activation code in the
text editor screen, which is opened after first attempt to select this function. Once unlocked this option
will be ready to use permanently.
Note: It is not possible to change the current function during running measurements.
In this case the instrument displays for about 2
seconds the warning: “Measurement in Progress”. To
change the current measurement function, the measurement must be stopped!
12.2 S
ETTINGS OF THE
RT60
ANALYSIS
Execution of the
RT60
analysis depends on a certain number of the parameters,
which can be set in the different screens of the
Measurement
section:
RT60 Settings
,
Compensation
Filter
and
Range
.
Positions
Compensation
Filter
and
Range
are the same as for other
instrument’s functions (see Chapters
The
RT60 Settings
list allows you to select
the method for
RT60
calculations, define the
name of the file, where the registered data
will be collected, and other parameters for
RT60
calculations.
The
Start Delay
parameter defines the
delay period from the moment the
<Start>
key is pressed to the measurement start.
<ENT>
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...