![Svantek SVAN 979 Скачать руководство пользователя страница 150](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285150.webp)
150
SVAN 979 User Manual
Max
For sound measurements: maximal value of the time-
weighted sound pressure level at the exponential RMS
detector output within the elapsed measurement time.
The
Max
result for the 1 second period is equal to the Spl
result.
For vibration measurements: maximal value of the time-
weighted
vibration
(acceleration,
velocity
or
displacement) level at the exponential RMS detector
output within the elapsed measurement time.
Appendix D
D.1.2,
D.2.2
Min
Minimal value of the time-weighted sound pressure level
at the exponential RMS detector output within the
elapsed measurement time.
Appendix D
D.1.2
Leq
Equivalent continuous sound level, time-averaged sound
level for the elapsed measurement time (equivalent
sound level).
Appendix D
D.1.2
LRm
Rolling (sliding)
Leq
screen for the last m minutes of
measurement (m x 60 seconds) moving with 1 second
step.
Chapter
Appendix D
D.1.2
SEL
Sound Exposure Level, the constant sound level that has
the same amount of energy in one second as the original
noise event and is the subset of the
Leq
result so, for the
integration time equal to 1 s,
SEL
is always equal to
Leq
.
Appendix D
D.1.2
LEPd
Daily Personal Noise Exposure, the noise exposure level
for a nominal 8-hour working day, used for assessing the
noise exposure of a worker during a working day. The
LEPd
result is calculated on the base of the
Leq
.
Appendix D
D.1.2
LTeq
Takt-Maximal Level calculated according to the German
standard TA Lärm. Applied filter and detector constant
are shown also in the name of this result with two letters,
as shown on the screen (A filter and F (Fast) detector).
Appendix D
D.1.2
Lnn
Statistical Noise Levels, the certain boundary level
surpassed by the temporary noise level values in not
more than nn% of the observation period.
Lnn
are
calculated on the base of 100ms Leq results and
renewed every second on the display as cumulated
statistics over the current measurement time.
Appendix D
D.1.2, D.4
L(den)
Day-evening-night equivalent level,
Leq
. Sound Level,
measured over the 24 hour period, with a 10 dB penalty
added to the levels between 23.00 and 07.00 hours and
a 5 dB penalty added to the levels between 19.00 and
23.00 hours to reflect people's extra sensitivity to noise
during the night and the evening. The instrument
displays: Ld, Le, Ln, Lde, Len, Lnd, or Lden depending
on the day and night time which the measurement
Appendix D
D.1.2
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...