![Svantek SVAN 979 Скачать руководство пользователя страница 193](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285193.webp)
193
SVAN 979 User Manual - Appendixes
1
– Turn off command
#R,0;
Response:
#R,0;
Note:
If any additional characters appear in the command after the number 0 (zero), they will be ignored, and
the command will not be treated as incorrect.
2
– Turn on command
#R,1[:[f][:[r][:[d]]]],C
1
[,C
2
[..[,C
n
]]];
where:
[]
– square brackets mean that the parameter is optional,
1
– a number indicating the activation of the 1-second results sending mode,
f,r,d
– transferred data format parameters
C
k
– results or result parameter identifier (k = 1..n)
The parameters of data transmission format have following meanings:
f
– line separator parameter – allows to send an additional character or line separator
characters after each transmission:
f
= 0
– no line separator sent (default)
f
= 1
–
line feed
character sent
(LF, ASCII value 10),
f
= 2
–
carriage return
character sent
(CR, ASCII value 13),
f
= 3
–
carriage return
i
line feed
character sent,
r
– results separator parameter – allows for change of results separator character (comma
character as default, ASCII value 44):
r
– any ASCII character from range 1 to 255
d
– decimal separator parameter – allows for change of decimal separator character (dot
character as default, ASCII value 46):
d
– any ASCII character from range 1 to 255
Note:
The lack of any data transmission format parameter between the colon characters will keep the default
value of the parameter. For example, command:
#R,1::9:44,(...);
will change the result separator to the horizontal tab (HT, ASCII value 9), change the decimal point to the
decimal point (ASCII value 44) and will not change the default line separator parameter.
Note:
If the results separator and decimal separator characters are identical, the command will not be
accepted, and the command error message will be returned - see point 4.
Each
C
k
identifier consists of a letter and possible numeric parameters separated by a colon. If the identifier is
given without the required parameters, default values of missing parameters will be zero. If the identifier will
contain more parameters than required, the command will not be accepted, and the command error message
will be returned - see point 4.
Available identifiers:
F[:p
1
]
– measurement start date; date is sent in a format depending on the parameter:
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...