![Svantek SVAN 979 Скачать руководство пользователя страница 160](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285160.webp)
160
SVAN 979 User Manual
Slope -
Type of trigger that starts the wave recording for
Recording Time
on condition: falling value of the
Leq/RMS
result
(
Source
)
integrated
during
Trigger Period
passes below the threshold value (Level).
Chapter
Level +
Type of trigger that starts the wave recording for
Recording Time
on condition: value of the Leq/RMS
result (
Source
) integrated during
Trigger Period
is
greater than the threshold value (Level).
Chapter
Level -
Type of trigger that starts the wave recording for
Recording Time
on condition: value of the Leq/RMS
result (
Source
) integrated during
Trigger Period
is lower
than the threshold value (Level).
Chapter
External
Type of trigger that starts the wave recording for
Recording Time
when the triggering signal appears on
the I/O socket of the instrument.
Chapter
Integr. Period
Type of trigger that starts the wave recording for
Recording Time
every time the measurement starts. If
Integration Period
is shorter
than
Recording Time,
the
wave recording will be continued for additional
Recording Time
.
Chapter
Source
Measured result that is compared with the threshold level
for triggering (
Level
) measured in the first profile: Leq(1)
in the sound mode and RMS(1) in the vibration mode.
Chapter
Level
Threshold level of the trigger condition.
Chapter
Trigger Period
Time interval of checking the triggering conditions. This
parameter can be set as: Log. Step (same as
Logger Step
value), 0.5ms, 100.0ms and 1s.
Chapter
Recording Time
Time of signal recording after meeting every trigger
condition. The available values can be selected from 1s
to 8h or Infinite. Recording stops after
Recording Time
or
earlier if the measurement is stopped manually.
Chapter
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...