When there is a connection to the USB interface (
USB Device
socket is
connected by the cable to a PC), the
„USB”icon is presented on the top of the
display and in the
Battery
screen there is the message:
USB Power
Voltage: x.xxV
.
When there is a connection to the
7
–16V
socket the
“plug” icon is presented on
the top of the display and in the
Battery
screen there is the message:
External Power Voltage: yy.yyV
.
2.7 D
EFAULT SETTINGS
Factory settings
The instrument as sold has default settings which you may change, but always
return to them with the use of
Factory Settings
option in the
Auxiliary Setup
menu.
Next chapters of the manual will describe in detail what each parameter means
and how to change the instrument settings.
Main default settings
With the default settings, the instrument will measure sound pressure level by
three virtual meters, so called profiles (Measurement
Mode: Sound
;
Measurement Function: Level Meter
) with 1 second delay from the
<Start>
key pressure (
Start Delay: 1 s
), 1 second integration time (
Integration Period:
00:00:01
), infinitive repetition till press the
<Stop>
key (
Repetition Cycle: Inf
),
linear integration for the RMS base results (
RMS Integration: Lin
), linear
integration for statistics (
Statistics: Lin
), daytime limits for Lden calculations
(
Day Time Limits: 6-18 h
), time frames for two rolling Leq calculations
(
Rolling Time(1): 30 m
and
Rolling Time2: 1 h
), free field compensation
(
Compensation Filter: Free Field
), active logging of the selected results with
1 second step (
Logger: On
;
Logger Step: 1 s
;
Logger Results: Peak
,
Max
,
Min
and
Leq
for all profiles) and summary results saving including
Statistics
.
The logger and summary results will be automatically saved in the file with the
name defined by the instrument and presented in the
Logger Setup
screen
(
Logger Name: Lxxxx
). Other functions are switched off, like:
- measurement trigger (
Measurement Trigger: Off
),
- logger trigger (
Logger Trigger: Off
),
- wave recording (
Wave Rec.: Off
),
- timer (
Timer: Mode: Off
).
Default profile settings for Sound measurements
:
Profile 1
-
A
weighting
filter
(
Filter(1)=A
),
Fast
RMS
detector
(
Detector(1)=Fast
);
Profile 2
-
C
weighting
filter
(
Filter(2)=C
),
Fast
RMS
detector
(
Detector(2)=Fast
);
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...