30
SVAN 979 User Manual
The screens below show the automatic file saving during two subsequent measurements. Before and during
the first measurement the file name
L30
is displayed. This file is saved automatically in the directory
hh-mm-
ss
after the measurement stop. After start of the second measurement the instrument automatically changes
the file name to
L31
and this file will be saved after of the second measurement stop and so on.
<Start>
<Stop>
<Start>
Note: Du
ring the measurement run with data logging to the logger file, the „logging” icon is
displayed.
If both
Logger
and
Summary Results
are
switched on, pressing the
<Save>
key
switches you to the
Logger Setup
screen in
which the
Logger Name
position shows the
name of the logger file which will be created
after the next measurement. You can
change parameters in this screen and return
to the measurement view after pressing
<ENTER>
.
<Save>
Manual saving of Summary Results
If the
Logger
or the
Summary Results
position is switched off in the
Logger Setup
list (
path: <Menu> / Measurement / Logging
/ Logger Setup
) the automatic saving of the
measurement
data
(so
called
Summary Results) is switched off too. In this
case Summary Results for the last cycle can
be saved only manually.
or
Note: If Logger is Off, the field with file name in the result view screen
will be empty!
There are two options for saving manually the
Summary Results
data. One option is to press
<Save>
key
(
<Alt>+<ENTER>
)
right after the measurement stop. Another option is to create
<New File>
with
File Manager
.
Pressing the
<Save>
key switches you to
the
New File
editor screen with the name
generated by the instrument.
After edition of the automatically proposed
name, press
<ENTER>
.
<Save>
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...