![Svantek SVAN 979 Скачать руководство пользователя страница 299](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285299.webp)
299
SVAN 979 User Manual - Appendixes
Note: For the signals with the crest factor n >1.41 upper measurement range of the RMS is reduced.
The valid upper limit can be calculated according to the below given formula:
, where A is the upper limit for the sinusoidal signal
Example: For the crest factor n = 10 the upper limit is A
10
= 120 dB
Measurement frequency range
with the
HP
filter (-3 dB)
0.4
Hz ÷ 22.4 kHz.
Maximum peak voltage
30 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be applied to the SLM without destruction the
meter)
RMS detector
•
Digital
“True RMS“ with Peak detection
•
Resolution
0.1 dB
•
Range
327.7 dB
•
Crest Factor
unlimited (for signals in 20 kHz band)
Reference conditions
•
Reference frequency
1000 Hz
•
Reference level
114 dB
•
Reference temperature
+20°C
•
Reference relative humidity
65%
Calibration (electrical)
Calibration level
114.0
dB (re 1μV
RMS
)
Basic accuracy
<
0.2
dB (for the temperature T=+23°C
5°C for sinusoidal
signal 1 V
RMS
in the band 10
Hz ÷ 20 kHz with
Z
filter)
Measurement error
in the full temperature range
<
0.1 dB (when the temperature is from -
10°C to +50°C for the
sinusoidal signal 1 V
RMS
in the band 10
Hz ÷ 20 kHz with
Z
filter)
Overload detector
The instrument has the built-in overload detectors. The overload in the measurement channel (in its analogue
part) and the overload of the analogue /
digital converter are both detected. The “overload” indication is when
the input signal amplitude is 0.5
dB above the declared “Peak measurement range”
Warm-up time
1 min. (for 0.1 dB accuracy).
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C re Reference conditions)
Effect of magnetic field
< 15 dB (A) or < 25 dB (Z) (for 80 A/m and 50 Hz)
Effect of Vibration
< 0.1 dB (from 20 Hz to 1000 Hz at 1 m/s
2
)
(
)
2
log
20
137
n
A
n
−
=
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...