Direct
– when this option is selected, the measured signal from the measurement channel is fed to
the pin 2 of the
I/O
connector. Output voltage, frequency band and the output impedance are
following:
a) Output Voltage:
1)
In the
Vibration Meter
and
Direct Vibration
modes, the output voltage is equal to 1.0 V
RMS
(
5 %) at 177 dB indication of the instrument, on
High
measurement range and at 160 dB
indication of the instrument on
Low
measurement range, when calibration factor is set to
0.0 dB.
2)
In the
Sound Meter
mode, the output voltage is equal to 70.7 mV
RMS
(
5 %) at 114 dB
reference level of the instrument, when the calibration factor is set to 0.0 dB.
3)
In the
Direct Sound
mode, the output voltage is equal to 1.0 V
RMS
(
5 %) at 137 dB
indication of the instrument, on
High
measurement range and at 120 dB indication of the
instrument on
Low
measurement range, when calibration factor is set to 0,0 dB.
b) Frequency Band (-3 dB):
1.0 Hz
÷ 22.6 kHz.
c) Output Impedance:
51
Ω / 5%
1.2.
D/A
– when this option is selected, the analogue signal from the internal signal generator of the
instrument after digital-to-analogue converter is transmitted to the pin 1 of the
I/O
connector, and the
analogue input signal is transmitted to the pin 2 of the
I/O
connector after digital-to-analogue
converter.
2.
I/O Mode: Digital In
. When the
Ext. Trigger
function is active, the external triggering of the instrument
may be provided. To do this, select
Trigger: External
(
path: <Menu> / Measurement / Measurement
Trigger
). The external signal for triggering is specified as follows:
a)
Trigger voltage threshold level is set to +1 V.
b)
Trigger voltage slope (
path: <Menu> / Measurement / Measurement Trigger
) set by the user as
Slope [+]
(uprising as default) or
Slope [
–]
(falling, auxiliary).
c)
Minimal duration of the trigger impulse: 10
μsec.
d)
100
μsec. release time after previous measurement is necessary before next trigger.
e)
Recommended trigger voltage should not exceed ± 5 V.
f)
Input impedance in the
Digital In
mode
– ca. 10 kΩ / 100 pF, ESD type safety.
g)
When the instrument is switched-off in the
Digital In
mode, the voltage impulse on the pin [2]
will
be able to switch-on the instrument, however in this case the minimal duration of the trigger
impulse of 100 msec is necessary, with uprising voltage slope.
3.
I/O Mode: Digital Out
. Two different functions are available in this mode:
3.1.
Function: Trigger Pulse
. When this function is selected, the terminal [2] is set as output, which
enables one to trigger another instrument (one instrument or more with trigger inputs connected
together in parallel), output trigger impulse meets specification given below:
a)
trigger impulse is generated before every measurement,
b)
output voltage range from 0 V to 3 V,
c)
triggering slope: uprising,
d)
output impedance: 51
Ω,
e)
duration of the impulse: ca. 40
μsec.
3.2.
Function: Alarm Pulse
. When this function is selected, the terminal [2] is set as an output,
which changes its output level, when current result of measurement exceeds user-
programmable threshold level. In this case the terminal [2] output operates as an output of
analogue comparator with user-programmable threshold. This feature enables one to control an
external device as alarm-indicator or similar:
a)
electrical specification of this output are as follows: 0 V to 3 V voltage range, 51
Ω output
impedance,
b)
output produces a voltage level (not impulse),
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...