![Svantek SVAN 979 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285037.webp)
37
SVAN 979 User Manual
2. Set the
Sensitivity
of the microphone taken from its calibration certificate using the
◄ / ► key and then
press
<ENTER>.
The calibration factor (
Cal. Factor
) is calculated every time after pressing the
◄ / ► key,
in relation to the nominal value of 50.00 mV/Pa.
For microphones with sensitivity higher than 50.0 mV/Pa the calibration factor
will always be negative.
For microphones with sensitivity lower than 50.0 mV/Pa the calibration factor will
always be positive.
The lowest available value of the sensitivity that can be introduced is equal to
50.0
µV/Pa (it conforms to the calibration factor equal to 60.0 dB) and the highest
value is equal to 50.0 V/Pa (calibration factor is equal to -60.0 dB).
3. To confirm the new calibration factor, press
<ENTER>
, or to return to the
Calibration
list without changes,
press
<ESC>
.
3.3.3
Calibration by Sensitivity (Vibration signal)
The calibration by using the accelerometer’s published sensitivity information
can be performed in the following way:
1. Select the
Calibr. by Sensitivity
position in the
Calibration
list and press
<ENTER>
.
2. Set the
Sensitivity
of the accelerometer taken from its calibration certificate
using the
◄ / ► key.
The calibration factor (
Cal. Factor
) is calculated every time after pressing the
◄ / ► key, in relation to the nominal value of 10.0 mV/ms
-2
.
For accelerometers with sensitivity higher than 10.0 mV/ms
-2
the calibration
factor will always be negative.
For accelerometers with sensitivity lower than 10.0 mV/ms
-2
the calibration factor
will always be positive.
The lowest available value of the sensitivity that can be introduced is equal to
10.0
µV/ms
-2
(it conforms to the calibration factor equal to 60.0 dB) and the
highest value is equal to 10.0 V/ms
-2
(calibration factor is equal to -60.0 dB).
3. To confirm the new calibration factor, press
<ENTER>
, or to return to the
Calibration
list without changes,
press
<ESC>
.
3.3.4
Calibration by Measurement (Acoustic signal)
Calibration by measurement of the acoustic signal can be done in the following
way:
1. Select the
Calibr. by Measurement
position in the
Calibration
list and press
<ENTER>
.
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...