38
SVAN 979 User Manual
2. Set the reference calibration level (
Cal. Level
)
– see Appendix C,
Chapter C.1, par. Calibration.
3. Attach the acoustic calibrator SV 36
(or equivalent
114 dB@1000 Hz)
carefully over the microphone of the instrument.
Note: It is also possible to use an electro-mechanical pistonphone, which generates a signal (ca
124
dB) or different type of acoustic calibrator dedicated for ½” microphones. It is also necessary to
switch the instrument Range to the High level.
4.
Switch on the calibrator (if the used calibrator doesn’t have auto run function)
and wait approximately 30 seconds for the tone to stabilise before starting
the calibration measurement.
5. Start the calibration measurement by pressing the
<Start>
key.
The instrument performs series of 1-second consecutive measurements
displaying levels of the measured calibration signal.
If maximal difference between three 1-second consecutive results (Leq(C)) is
less than
0.05dB
, the calibration measurement will be stopped, and the
calibration factor calculated. The measurement can be always stopped by the
<Start/Stop>
key.
The
Calibration Result
(measured reference signal without calibration factor
correction) and the
New Calibration Factor
(difference between the Calibration
Level and the Calibration Result, calculated in dB) are displayed.
6. Press
<ENTER>
to save the new calibration factor (
Enter to Confirm
), or
press
<ESC>
to reject it.
7. Press
<ENTER>
to exits the
Calibration by Measuement
screen.
It is recommended to repeat the calibration measurement few times. Obtained results should be almost the
same (with
0.1 dB difference). Reasons for unstable results are as follows:
•
the calibrator is not properly attached to the instrument,
•
there are external acoustic disturbances such as high noise levels close by,
•
the calibrator or the measurement channel (the microphone, the preamplifier or the instrument itself) are
damaged.
8. Detach the calibrator from the microphone.
Note: During the calibration measurement, any external disturbances (acoustic noise or vibrations)
should not exceed a value of 100 dB (when using a calibrator that generates a level of 114 dB).
3.3.5
Calibration by Measurement (Vibration signal)
Calibration by measurement of the vibration signal can be done in the following
way:
1.
Select the
Calibr. by Measurement
position in the
Calibration
list and
press
<ENTER>
.
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...