![Svantek SVAN 979 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285043.webp)
43
SVAN 979 User Manual
4.1 S
ETTING UP GENERAL MEASUREMENT PARAMETERS
–
G
ENERAL
S
ETTINGS
The
General Settings
list allows you to set
up general measurement parameters:
measurement start delay (
Start Delay
),
maximum
delay
period
for
the
synchronization with RTC (
Start Sync.
),
measurement period (
Integration Period
),
repetition
of
measurement
cycles
(
Repetition Cycles
), RMS detector type
(
RMS Integration
), RMS detector type for
statistic results (
Statistics
), day time hours
(
Day Time Limits
) and time window for the
LRm
results (
Rolling Time
).
▼
Delay of measurement start
The
Start Delay
parameter defines the delay period from the
<Start>
keystroke
to the real start of the measurement (digital filters of the instrument constantly
analyses the input signal even when the measurement is stopped). This delay
period can be set from
0 second
to
60 seconds
.
Synchronisation of measurement start
The
Start Sync.
parameter defines synchronisation points with the instrument’s
RTC and can be set as:
Off
,
1m
,
15m
,
30m
and
1h
. For example, if
1h
is
selected, the measurement will start from the beginning of the first second of
next hour after the
<Start>
keystroke, and then will be repeated also from the
first second of the following hour after elapsing the integration period if the
number of cycles is greater than one. Default value:
Off
.
Disabling the measurement period definition
The integration period can be set as infinite or can be defined together with the
Repetition Cycles
number. The
Integr. Period Inf
parameter activates the
infinite integration period. In this case the signal will be averaged all the time until
the
<Stop>
keystroke.
If the
Integr. Period Inf
parameter is On, the
Split Cycles
position appears.
The
Split Cycles
position enables you to limit the number of logger files when
the measurement period is not defined. If you select the number instead of
Inf
the logger splitting will be performed no more times than this number.
If the
Integr. Period Inf
parameter is Off, instead the
Split Cycles
position two additional positions appear:
Integration Period
and
Repetition Cycles
.
Measurement period
The
Integr. Period
parameter (integration
period) defines the period during which the
signal is being measured (and for some
results integrated) and measured values
logged
in
the
result
file
as
Summary Results
. The integration period
can be set in the special screen, which is
opened with the
◄ / ► key.
►
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...