![Svantek SVAN 979 Скачать руководство пользователя страница 367](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285367.webp)
367
SVAN 979 User Manual - Appendixes
The measurement time in this method consists of:
•
The time between markers t
on
and t
off
in which the omnidirectional sound source emits acoustic power and
the SVAN xxx analyser measures the actual sound pressure level.
•
The time between markers t
off
and t
trig
in which the omnidirectional sound source is switched off and the
SVAN xxx instrument waits for trigger condition fulfilment.
•
The time between markers t
s
and t
trig
registered since the trigger condition fulfilment back till point t
s
to allow
recognising the beginning of the decay region. In the SVAN xxx instruments this time is equal to
Time Step
(
path: <Menu> / Measurement / RT60 Settings
) parameter value multiplied by 50.
•
The time between markers t
trig
and t
e
registered since t
trig
forward to record whole decay curve together with
significantly long period of the noise level. This time in SVAN xxx instruments is adjusted by
Recording Time
(
path: <Menu> / Measurement / RT60 Settings
) parameter.
The above graph shows that the proper setting of the
Recording Time
value is very important. The registration
time has to be long enough to acquire sufficient number of background noise level values. In other case the
decay curve region could not be properly analysed or decay region could not fulfil the dynamic condition
mentioned above. It is recommended to set the
Recording Time
parameter two times longer than expected
reverberation time.
➢
IMPULSE method
In the Impulse method, Reverberation Time is computed by using the reverse-time integrated impulse
response. This way of measuring sound decay was introduced firstly by M. R. Schroeder in two historical
articles:
o
New Method of Measuring Reverberation Time,
Journal of Acoust. Soc. Am.1965
o
Integrated-Impulse Method Measuring Sound Decay without Using Impulses,
Journal of Acoust. Soc. Am.
Vol. 66(2) 1979
Fig. 6 An example of Schroeder integration with the limits Ti and Td
This RT 60 measurement method requires impulse sound source like pistol, petard or other sound source
which emits impulse signal with very high sound pressure level.
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...