SVAN 979 User Manual
The calibration record screen contains the
information
regarding
the
performed
calibration: date and time, used calibration
method, obtained calibration factor.
<ENT>
3.3.7
Erasing calibration records
– Clear Calibr. History
You can erase all calibration records. To do
this, choose the
Clear Calibr. History
position in the
Calibration
list and press
<ENTER>
to perform this operation.
The instrument will request the confirmation
of this operation.
<ENT>
After erasing calibration records, the
Calibration History
screen will not contain
any previous calibration records. The content of this screen is also cleared after
the
Factory Settings
operation (
path: <Menu> / Auxiliary Setup
).
Note: Acoustic and vibration calibration histories are independent files in
the internal instrument’s
memory and erasing the calibration history in one mode does not erase it in another mode.
3.3.8
Automatic calibration
– Auto Calibration
Automatic
calibration
feature
enables
calibration of the instrument input as easy as
possible with minimum steps.
If the automatic calibration is switched on,
the instrument, when it doesn’t perform the
measurement, periodically compares the
measured signal level with the reference
calibration level and starts the calibration
measurement if the stable result is within
±5dB of the calibration level.
►
To perform the automatic microphone calibration, follow next steps:
1. Switch on the instrument.
2. Attach the SV 36 (or equivalent 114 dB@1000 Hz) calibrator to the microphone and switch it on (if the
used calibrator doesn’t have automatic switch-on feature).
Generated by the calibrator sound pressure level starts the Automatic
Calibration process if the difference between the
Calibration
Level
value set
up in the
Calibration By Measurement
screen and the measured calibrator
SPL level is in the range
±5dB
.
During the calibration measurement, the level of the measured calibration signal
will be displayed as in case of calibration by measurement The measurement
can be always stopped by the
<Stop>
key.
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...