SVAN 979 User Manual
10
1/1-OCTAVE
– 1/12-OCTAVE ANALYSER
The instrument operates as a real time 1/1-octave - 1/12-octave
analyser (RTA) in a very similar way to the
level meter. Moreover, 1/1-octave - 1/12-octave
analysis is performed in parallel with the level meter
measurements. All 1/1-octave, 1/3-octave, 1/6-octave and 1/12-octave digital pass-band filters are working in
real-time with the weighting filters (
HP
,
Z
,
A
,
B
or
C
-
for sound measurements;
HP
- for vibration
measurements) and the RMS/LEQ detector (
Linear
,
Fast
or
Slow
-
for sound measurements;
Linear
- for
vibration measurements). This enables a spectrum pre-weighting with one of the selected broadband frequency
curves if required for the application such as the provision of hearing protectors during the control of high
workplace noise.
Note: TOTAL RMS (LEQ) results are measured with their own weighting filters (A, C, Z or S1, S2,
S3 - for sound measurements; HP, HP3, HP10 or S1, S2, S3
– for vibration measurements)
regardless of settings made in profiles for the Level Meter measurements. Spectra are always
linearly averaged. Thus, the Total values for 1/1-octave - 1/12-octave analysis
can be different from
those obtained for profiles (if RMS Integration was set as Exponential).
For each 1/1-octave - 1/12-octave
band, the
Leq
/
RMS
result is calculated and presented
as a bar on the spectrum plot. Spectra can
be examined on a display in the
Spectrum
view.
The read-out of spectrum values can be
done using a vertical cursor.
Besides results for bands three
Total
values
are measured and displayed as additional
three bars on the spectrum plot. Parameters
for
Total
values (e.g. filters) are set by
default and cannot be changed.
Besides three
Total
values you can define
up to three additional
User Total
values with
filters:
S1
,
S2
,
S3
,
S4
and
S5
.
10.1 S
ELECTING THE
1/1 O
CTAVE
–
1/12 O
CTAVE FUNCTION
To select the 1/1-octave - 1/12-octave
analysis
function,
open
the
Measurement Function
screen, select the
1/1 Octave
-
1/12 Octave
position
and
press
<ENTER>
.
<ENT>
Note: The 1/6 Octave and 1/12 Octave functions are optional and should be unlocked by entering
an activation code in the text editor screen, which is opened after first attempt to select one of this
option. Once unlocked the option will be ready to use permanently.
Note: It is not possible to change the current function during a running measurement.
In this case, the instrument displays for about 2 seconds the message:
“Measurement in Progress”.
To change the current function, the measurement must be stopped!
10.2 C
ONFIGURING
1/1-
OCTAVE
–
1/12-
OCTAVE ANALYSER
Execution of the 1/1-octave - 1/12-octave
analysis depends on a certain set of parameters, configured in the
Measurement
section.
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...