![Svantek SVAN 979 Скачать руководство пользователя страница 264](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285264.webp)
264
SVAN 979 User Manual - Appendixes
Preamplifier
SV 17
nominal preamplifier attenuation: 0.7 dB;
Voltage Type
– power supply 1.5 mA@30 V
7 pin Lemo connector. See chapter C.3 for the
SV17 connector pin-out.
Maximum peak voltage
30 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be lead to the SLM without destruction the
meter)
Warm-up time
1 min. (for 0.1 dB accuracy)
Nominal delay
less than 3 seconds
(between operating of the "Reset- Button"
and beginning of a new measurement)
Typical stabilization time
after change in
environmental conditions by 20
C
1 hour
Time shift after completion
of a
measurement, before a measurement is
shown
< 1 sec
Note: When the instruments are moved from a warm environment with high humidity,
to a colder environment, care should be taken not to produce condensation inside
the instruments. In this case, much longer stabilization periods may be necessary.
Environmental, electrostatic
and radio frequency
criteria
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C and 1000 Hz)
Effect of magnetic field
< 15 dB (A) or < 25 dB (Z) (for 80 A/m and 50 Hz)
Effect of radio frequency fields
< +/-0.5 dB @ 74 dB and 10V/m
electromagnetic field
The greatest susceptibility (the least immunity) is achieved when in the SLM the
Z
filter and time weighting
F
are selected, and the SPL measurements are considered.
The greatest susceptibility is achieved when the SLM is placed parallel to the radio frequency field.
In addition, if there is an extension cable, the greatest susceptibility is achieved when the SLM and cable is
placed along field and the cable is coil as solenoid.
Effect of electrostatic discharge
meets requirements of IEC 61672-1:2013
During electrostatic discharge, the influence of the displayed results could be observed.
No changes in instrument operation state, configuration or stored data corruption were found out.
Effect of ambient pressure
< 0.01 dB/kPa
Effect of temperature
< 0.5 dB (from -10
C to + 50
C)
Operating range
from -10
C to + 50
C
Storage
from -20
C to + 60
C
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...