![Svantek SVAN 979 Скачать руководство пользователя страница 292](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285292.webp)
292
SVAN 979 User Manual - Appendixes
C.2.
S
PECIFICATION OF
SVAN
979
AS A
V
IBRATION
L
EVEL
M
ETER
The configuration of the complete VLM
and its normal mode of operation
SVAN 979
sound & vibration analyser
SV 80
10mV/m/s2, IEPE accelerometer (see C.6 for details)
SC 27
coiled cable for accelerometer 2 m
SV 18T
LEMO 7-pin to TNC integrated connector
Accessories included
in SVAN 979 instrument set
SC 59
Output cable for I/O connector, LEMO to 2 x TNC
SC 16
USB 1.1 cable,
SA 33
Power supply unit.
Accessories available
SV 55
RS 232 interface for SVAN 95x instrument series
SC 18T
LEMO 7-pin to TNC integrated connector
SC 27
TNC (plug) to TNC (plug) 2 m coil cable
SA 17A
external battery pack
SA 27
permanent magnet
SA 15
power supply unit
SV 110
vibration calibrator (or equivalent)
Measured quantities
The measured quantities in the vibration meter mode are:
RMS
,
PEAK
,
P
–P
,
MAX
and
MTVV
.
Definitions for mentioned parameters are given in Appendix D.
Mounting for vibration tests
The accelerometer can be connected with the VLM using proper cable provided by the manufacturer.
The accelerometer can be mounted on the plate in various ways:
•
using threaded stud onto a flat, smooth surface,
•
using thin layer of bees-wax for sticking the accelerometer into the plate,
•
using mica washer and isolates stud, where the body of accelerometer should be electrically isolated from
the measuring object,
•
using permanent magnet.
Values of the measured acceleration
Linear operating ranges for the acceleration
(with the SV 80 sensitivity equal to 10mV/ms
-2
accelerometer.
Two ranges with the nominal values:
1
00 ms
-2
(
LOW
), (140 dB, related to 10
-6
ms
-2
).
352 ms
-2
(
HIGH
) (151 dB related to 10
-6
ms
-2
).
•
Linear measurement with the
Wh
filter:
from 0.003 ms
-2
to 352 ms
-2
(the sinusoidal signal RMS)
from 0.03 ms
-2
to 500 ms
-2
(PEAK)
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...