![Svantek SVAN 979 Скачать руководство пользователя страница 183](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285183.webp)
183
SVAN 979 User Manual - Appendixes
#1,Xccc,X?,Xccc,(...),X?,Xccc;
where:
X
- the group code,
ccc
- the code value,
X?
- the request to send the current X code setting.
The instrument outputs in this case a control settings file for all requests X? in the following format:
#1,Xccc,Xccc,(...),Xccc;
In order to read out all current control settings the user should send to the device the following sequence of
characters:
#1;
The instrument outputs in this case a file containing all control settings given in Tab. A1
in the format:
#1,Xccc,Xccc,(...),Xccc;
Note: All bytes of that transmission are ASCII characters.
17.8 A.2.
F
UNCTION
#2
–
R
EAD OUT OF THE MAIN MEASUREMENT RESULTS
A.2.1. Read-out the profile results
#2
function enables one to read out the current measurement result from the selected profile.
#2 function
has the format defined as follows:
#2,t,p,X?,X?,X?,(...),X?;
where:
t
– the type of the results (
a
– averaged
,
i
– instantaneous
,
c
– 1 second results
)
p
– the number of the profile (
1
,
2
or
3
) or
0
for all profiles.
X
– the code of the result,
The instrument sends the values of results in the format defined as follows:
#2,t,p,Xccc,Xccc,Xccc,(...),Xccc;
(where
t=a/i/c
,
p=1/2/3
)
or
#2,t,1,Xccc,(...),Xccc,2,Xccc,(...),Xccc,3,Xccc,(...),Xccc;
(when
t=a/i/c
,
p=0
)
or
#2,?;
(when the results are not available).
If the command contains status query (optional
s
letter) #2 function has the format as follows
#2,ts,p,X?,X?,X?,(...),X?;
The instrument sends the values of results in the format defined as follows:
#2,tsf,p,Xccc,Xccc,Xccc,(...),Xccc;
(where
t=a/i/c
,
p=1/2/3
)
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...