Slope trigger
The
Slope + / Slope -
trigger starts the measurement/integration with the
duration of the
Integration Period
on condition: rising value of the measured
result (
Source
) averaged by 0.5 ms passes above/below the threshold value
(
Level
).
When the new measurement cycle begins (after pressing the
<Start>
key or automatically after the stop of the
previous measurement cycle) the instrument checks the trigger condition every 0.5 ms and if condition is met
the instrument starts the continuous series of 1-second integrations, the number of which is equal to the number
of seconds in the
Integration Period
.
After ending the
Integration Period
, the
new measurement cycle can start with
above logic.
The measurement can be stopped manually
at any moment with the
<Stop>
key.
Note: When a measurement is waiting for the slope trigger, the
“slope” icon appears alternatively with the „measurement” icon.
or
Level trigger
The
Level + / Level -
trigger starts the 1-second measurement/integration on
condition: value of the measured result (
Source
) averaged by 0.5 ms is
greater/lower than the threshold value (
Level
). In other cases, the instrument
continues checking the trigger condition every 0.5 ms.
When a new measurement cycle begins (after pressing
<Start>
key or automatically after previous
measurement cycle stop) the instrument checks the trigger condition every 0.5 ms and starts 1-second
integration if condition is met.
After 1-second integration, the instrument
repeats the trigger condition checking every
0.5 ms and starts next 1-second integration
if condition is met. Instrument does it as
many times as many seconds are within the
Integration Period
and
stops
the
measurement cycle. Therefore, the series of
1-second measurements may not be
continuous, and the duration of the
measurement cycle may be longer than the
Integration Period.
The measurement can be stopped manually at any moment with the
<Stop>
key. Summary Results are
calculated on the base of series of 1-second results measured during each measurement cycle and saved in a
logger file.
Note:
When a measurement is waiting for the level trigger the “level”
icon appears alternatively with the „measurement” icon.
or
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...