![Svantek SVAN 979 Скачать руководство пользователя страница 156](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285156.webp)
156
SVAN 979 User Manual
External
Type of trigger that starts the measurement/integration
when the trigger signal appears on the pin 2 of the I/O
socket of the instrument. After the measurement/
integration start from the trigger, the measurement/
integration will continue by the
Integration Period
.
Chapter
Source
Measured result that is compared with the threshold
value (Level) for triggering
– RMS measured in the first
profile: Leq(1) in sound modes and RMS(1) in vibration
modes.
Chapter
Level
Threshold level of
Source
for triggering condition
fulfilment.
Chapter
Gradient
Threshold level of the Source gradient for triggering
condition fulfilment.
Chapter
RPM
Measurement
Screen that enables configuring measurement of the
rotation per minute/second. To perform the RPM
measurement the RPM probe should be connected to the
I/O socket.
Chapter
Exposure Time
Screen that enables setting a duration of worker’s
exposition to the noise. This duration is considered in
LEPd calculations.
Chapter
Appendix D
D.1.2
Statistical Levels
Screen that enables setting a boundary level (
Lnn
)
surpassed by the temporary noise level values in not
more than nn
%
of the observation period. The user can
define ten statistical levels, named from N1 to N10, to be
calculated, displayed and saved in the files together with
the main results.
Chapter
Appendix D
D.4
Timer
Screen that enables configuring automatic switching
On
the instrument and performing the measurement on the
programmed time with defined setup. Timer can be
Single
or repeatable (
Multiple
).
After every timer cycle, the instrument automatically
switches itself off.
Chapter
Logging
Screen that enables configuring saving of the
Summary Results
,
Logger Results,
markers and a
waveform signal in files with the use of the next screens:
Logger Setup, Logger Results, Summary Results,
Logger Trigger, Marker Setup
and
Wave Recording.
Chapter
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...