InnoVet Select
™
HF APR
Service Manual
56
9.0 GENERATOR CALIBRATION USING THE OPERATOR CONSOLE
Calibrations will be done through software adjustments on the Operator Console. The
XRAY CAL screen, which is password protected, allows access to the screens which
control kVp, mA leading edge, and stabilized mA levels.
NOTE: It is best to view both the kVp and the mA waveforms simultaneously
when calibrating the system. The PWR MOD CAL screen is set up to allow
control of mA leading edge, stabilized mA and stabilized kV all at one time.
9.1 Required Equipment for Calibration
Recommended test equipment is as follows:
•
Dual trace storage oscilloscope
•
mR meter (for verification of linearity and reproducibility)
•
Dynalyzer or equivalent
(If Dynalyzer is unavailable the following equipment can be used as a
replacement. See below for further explanation.)
•
Non-invasive kVp meter
•
Digital mAs meter
•
Digital
Multimeter