InnoVet Select
™
HF APR
Service Manual
18
1.10 Power Line Requirements
The line currents given in Table 1-2 (for the 208 to 240 volt line) are the momentary
line currents during an exposure. This is not the recommended circuit breaker
rating. The recommended supply power requirements are:
100AMP circuit breaker
37.5 kVA distribution transformer
#2 AWG wire for 50 feet run
#00 AWG wire for 100 feet run
250 mcm wire for 200 feet run
Maximum load voltage drop under full load to be not more than 5%
These wire sizes are for copper conductors between the distribution transformer
and the 100 Amp service disconnect. The 100 Amp breaker listed for 240 VAC
operation must be upgraded to 125 Amp for operation below 240 VAC as indicated
by the line currents listed in Table 1-2 above.
1.11 Preliminary Room Inspection
Prior to beginning installation, inspect the x-ray room to verify compliance with
specifications for the following:
•
Incoming Line
•
Service Disconnect
•
Lead Shielding per local code
•
Conduit
•
Space
Requirements
1.12 Unpacking
Upon receipt of the x-ray generator and associated equipment, inspect all shipping
containers for signs of damage. If damage is found, notify the carrier or his agent
immediately. When the equipment is unpacked, inspect all pieces for visible
damage. If any damaged parts are detected, repair or order replacements to
prevent unnecessary delay in installation.
Before beginning the installation, verify the following:
•
All internal connections and fasteners are secure
•
All internal wiring is secure