InnoVet Select
™
HF APR
Service Manual
48
Table 7-1: DIP1 Functions (System Controller PCBA)
Dip Switch
OPEN (OFF)
CLOSED (ON)
DIP1-1
Normal Operation
Simulator mode (demo)
DIP1-2
Normal Operation
Disables capacitor bank charging.
Exposure with mA and kVp is allowed to
verify feedback system. Exposures will
discharge the capacitor bank.
DIP1-3
Normal Operation
Rotor and filament disable: this allows
exposure (kVp only) to check for high
voltage breakdown of the transformer
(high pot test).
DIP1-4
mA feedback OFF
(mA feedback disabled)
Normal Operation
mA feedback ENABLED (ON)
DIP1-5
Normal Operation
Not used
DIP1-6
Normal Operation
Not used
DIP1-7
Normal Operation
Not used
DIP1-8
Normal Operation
When booted up in cal mode this reloads
the factory defaults for all configuration
and calibration data. Used to clear E14
Table 7-1A: DIP1 Functions (System Controller PCBA)
DIP
Switch
OPEN (OFF)
CLOSED (ON)
DIP3-1
Normal Operation
Sets mA feedback scale for 300mA
Sets mA feedback scale for
500mA option.
DIP3-2
Normal Operation
Sets mA feedback scale for 300mA
Sets mA feedback scale for
500mA option.
DIP3-3
Normal Operation
Not used
Not used
DIP3-4
Normal Operation
Used in service mode only, to
allow adjustment of steady state
exposure mode inverter
frequency for 500mA units.