Select 50 kVp, 200 milliseconds and 50 mA. Take exposures and adjust the mA
OFFSET value as needed to achieve the proper stabilized mA and mAs. SAVE the
change.
Using the same kVp and time, now select 300 mA. Take exposures and adjust mA
SLOPE as needed to achieve the proper stabilized mA and mAs. SAVE the
change.
6.
Verify mA outputs are square for all mA stations across the kVp range.
Take exposures at 40, 50, 70, 90, 110 and 125 kVp at all mA stations (where
allowed by kW limits) using 50 milliseconds. Adjust the individual FIL AMPS
number at these points as needed to fine tune the mA leading edge to achieve a
square waveform of the proper mAs.
7.
Calibration of kVp.
Select 70 kVp, and 50 mA. Select a time appropriate to your test equipment, not
less than 100 milliseconds. The MASTER kV adjustment will appear in the upper
right corner of the screen as shown below. Take exposures and adjust the kVp by
changing the MASTER kV value. One number change is equal to one kVp.
POWER MOD
kVp STEP:70 CALIBRATION MASTER kV:-2
FIL
AMPS:4.26
mA:50 SM
MASTER mA
TIME(ms):100 TABLE TOP SAVE
PAGE
POWER MOD
kVp STEP:50 CALIBRATION
FIL AMPS:4.09
mA 0
mA:50 SM
MASTER mA
TIME(ms):200 TABLE TOP SAVE
PAGE
POWER MOD
kVp STEP:50 CALIBRATION
FIL AMPS:4.83
mA SLOPE:+17
mA:300 LG
MASTER mA
TIME(ms):200 TABLE TOP SAVE
PAGE