InnoVet Select
™
HF APR
Service Manual
28
WARNING
:
Particular care should be taken when servicing the inside of the tubestand.
There is an extreme threat of mechanical pinching between the vertical slide
and counterweight due to their close proximity and opposite directions of
motion. All of the movable assemblies and parts of x-ray equipment should
be operated with care. Only properly trained and qualified personnel should
be permitted access to any internal parts.
4.
It is now safe to remove the shipping bolts which lock the main
counterweight and vertical carriage together, taking care to insure that the
counterbalance is adequate to avoid uncontrolled motion of the vertical slide.
Remove the back panel of the tubestand for access to all tubestand
adjustments. Any additional counterweights required to counterbalance the
tube arm vertically can be placed in the brackets at the top of the main
counterweight.