InnoVet Select
™
HF APR
Service Manual
45
6.10 Table Interface Connections TB5
NOTE: The TB5 terminal strip located in the power module is used to interface the
collimator, prep/exposure switch, air switches etc. to the generator via the table interface
cable.
Table 6-3: Table Interface Cable Connections at TB5
Origin Destination
Function
Collimator cable C1
TB5 - 12
24 VAC
Collimator cable C2
TB5 - 13
24 VAC
Collimator cable GND
TB5 - 11
GROUND
Collimator cable CSW1
TB5 - 5
Hip switch
Collimator cable CSW2
TB5 - 6
Hip switch
Table Interface cable - BROWN
TB5 - 2
Common
Table Interface cable - BLUE
TB5 - 4
Exposure
Table Interface cable - RED
TB5 - 5
Hip switch for collimator
Table Interface cable - GREEN
TB5 - 6
Hip switch for collimator
Table Interface cable - ORANGE
TB5 - 7
+ 24 VDC
Table Interface cable - VIOLET
TB5 - 8
To float-top lock switch
Table Interface cable - BLACK
TB5 - 9
To Solenoid
Table Interface cable - WHITE
TB5 - 10
- 24 VDC
(table continued on next page)