6.0 Generator Installation Procedure
6.1 Line
Voltage
Measure and note the incoming line voltage. 240 VAC is required for full 30 kW
output. If 30 kW output is desired and the line voltage is below 240 VAC, a line
matching transformer kit (part number L246) will be required.
NOTE:
If the line voltage is below 240 VAC (i.e., 208 VAC), it will not be
possible to produce the full output capability of the generator
without a line-matching transformer.
6.2 Position System Components
Place the Power Module close to the front of the table but not inside, leaving access
on all sides. This allows the installer to establish the proper wiring, routing and
lengths for all cables.
6.3 Operator’s Console
Run the two communication cables from the operator’s console to the Power
Module. (See section 2.5 for mounting the operator’s console and communication
cable routing.)
6.4 Transformer
Installation
Place the HV Transformer in front of the bottom shelf of the Power Module. The
high voltage cables cannot be inserted or removed when the transformer is inside
the table.
NOTE:
Release the 8-32 screw in the oil fill plug (1) full turn counter-
clockwise to allow the oil level to expand with temperature and
barometric changes.