InnoVet Select
™
HF APR
Service Manual
54
8.4 TUBE SETUP Screen
TUBE:
Press the arrow buttons beside the display to toggle through the compatible
tubes. Select the tube type being used here. This selection allows the
installer to choose between keeping the existing calibration data or loading
the factory default calibrations for this tube type.
kV LIMIT: Default is 125. Can be reduced if high voltage breakdown is a problem.
STND-BY I: Sets stand-by filament current, factory set to 2.5 filament Amps.
BST TIME: Rotor delay time, factory set to 1.5 seconds.
E-CNT:
Exposure counter. Automatically records total # of exposures taken.
Note: Press the SAVE PAGE button to store any changes to the values.
Press the BACK button to return to the XRAY CAL screen.
8.5 PWR MOD SET Screen
The PWR MOD SETUP screen will display as shown below.
KW LIMIT: Sets kW limit, adjustable in 1 kW increments. Values lower than 30 allow for
operation on poor power lines where line voltage drop is a problem.
Maximum mAs and mA will not be affected, only kW.
T RECP:
Toggles between BUCKY, GRID (grid cabinet), and OFF (no receptor).
W RECP:
Toggles between BUCKY, GRID (grid cabinet), and OFF (no receptor).
SHUT DOWN TIME: Selects hours of inactivity which cause automatic turn off of
control.
kV OVERSHOOT: Factory set only, 80 is default value.
kV SETTLE: This number is factory set to 10. It controls the kVp leading edge. Only in
rare cases will a change in this number be required. Contact the factory for
more details before making a change to this number.
Note: Press the SAVE PAGE button to store any changes to the values.
Press the BACK button to return to the XRAY CAL screen.
POWER MOD SHUT DOWN
KW LIMIT:30 SETUP TIME:2 hr
LINE FREQ:60
KV OVRSHT:80
T RECP:GRID
KV 10
W RECP:OFF