InnoVet Select
™
HF APR
Service Manual
31
3.4 Verification Of X-Ray To Image Receptor Alignment
1.
Any left-to-right adjustment required for coincidence between the central ray
and the center of the image receptor can be achieved by adjusting the grid
cabinet-to-tubestand interlock bracket, located on the lower front surface of
the tubestand.
2.
Any front-to-back adjustment of the central ray (after the arm has been
leveled from 3.1 above) can be achieved by loosening the tube arm collar
screws and sliding the tube arm shaft within the collar to achieve proper
alignment. Total adjustment is ¾ inches.
3.5 Securing Table To Floor
Once the table is in final position and all checks have been made, it can be secured
to the floor using the four (4) clamps provided with the feet. Slide the clamp into the
notch at the base of each foot, turn at 45
o
to the table and secure with a lag screw.