8.6 PWR MOD CAL, FIL# TBL Screens
These screens are branches of the XRAY CAL screen which relate specifically to
calibration of kVp and mA.
Their use will be described in
Section 9 - Generator Calibration
.
8.7 MORE Screen
This screen is reserved for factory use only.
Changing the factory defaults for FIL2
and FIL4 will put the x-ray tube at risk and void warranty. There are no values in
this screen that may be changed in the field.
XRAY CAL +
MISC.1 FIL 2 A:-0.14
FIL 4 A:-0.22
REV. LCD: N
E-CNT
RESET
SAVE
PAGE