InnoVet Select
™
HF APR
Service Manual
42
TB1
1
(256)
2
(240)
(220)
3
SEE NOTE 6
5
(120)
6
(0)
4
(208)
SEE NOTE 6
TB1
(240)
2
1
(256)
(220)
3
(208)
4
(120)
5
(0)
6
ALL LEADS LABELED 1 TO 6
ARE FROM LINE MATCH
X-FMR. THE LEADS #'S
ALL LEADS LABELED 1 TO 6
ARE FROM LINE MATCH
X-FMR. THE LEADS #'S
MATCH THE TERMINAL #'S.
INVERSE TO THE TERMINAL #'S.
6. On the left side of TB1, locate the wire labeled “1”. This is the line voltage adjust wire.
Move wire “1” to the TB1 tap which most closely matches the incoming line voltage
(208 VAC or 220 VAC).
7. On TB2, located just in front and to the right of TB1, verify the following connection:
The wire labeled “2J” should be on TB2-1, the terminal labeled “240”.
8. When all connections above have been verified, it is safe to turn on the mains input
voltage. Confirm that the voltages on TB1 and TB2 are correct as referenced to the “0
Volt” tap on the respective terminal block.
9. Once the voltages have been verified, turn off the mains input voltage and continue
with the installation and calibration procedure of the generator as outlined in the
manual.