InnoVet Select
™
HF APR
Service Manual
66
Error Code
Likely cause
How to clear the fault
Secondary
actions
ERROR
SYSTEM
COOL DWN
Wait 10 times
exposure length
before next exp.
Press any button.
Unit counts
down until
exposure is
allowed.
INVALID
APR DATA LIMIT
Receptor
configuration at
odds with APR
receptor selection
Press RESET and
correct configuration
error.
See
troubleshooting
guide below.
ERROR
DOOR
INTLK
Open connection
or opto chip on
01003-000pcb.
Verify TB3-5 is at
ground. Press any
button.
ERROR
TUBE INTLK
Open connection
or opto chip on
01003-000pcb.
Verify TB3-4 is at
ground. Press any
button.
ERROR
COLMTR INTLK
Open connection
or opto chip on
01003-000pcb.
Verify TB3-6 is at
ground. Press any
button.
ERROR
GENRL INTLK
Open connection
or opto chip on
01003-000pcb.
Verify TB3-7 is at
ground. Press any
button.
ERROR
DC BUS LOW
RESTART
SYSTEM
Checked at prep;
not enough
capacitor bank
voltage.
Cycle power and find
damaged component.
Confirm K1,
BR1, R1, K660
Relay pcb and
K445 Charge
Monitor pcb are
OK.
ERROR
STILL ANODE
No rotor current
during prep.
Press any button.
Check rotor
supply voltage,
stator.
ERROR
ANODE ROT
RESTART
SYSTEM
Rotor current
sensed when not
in prep.
Cycle power and find
defective
components.
Confirm rotor
voltage and
SSR1 is not
shorted.
ERROR
PREP HELD
Prep switch or
operator.
Release prep button.
20 sec is max
prep time.
ERROR
NO BUCKY MOT
No B1/B2 closure. Confirm closure and
opto. chip on 01003-
000, press any key.
Confirm image
receptor
configuration