11.3
The DEFAULT VET screen (Memory card not required)
The DEFAULT VET screen will display as shown below.
DFLT APR&CAL allows reloading the factory defaults for both calibration and APR data
into the Power Module and Operator Console. Because calibration data is involved,
this screen is password protected using the XRAY CAL password. The Console will
then ask for confirmation with a NO or YES answer before replacing the data.
DFLT ALL APR allows the user or service personnel to replace all of the existing APR
techniques with the factory default values. No password is required, just a NO or YES
response.
DFLT APR REG allows the user or service personnel to replace the existing APR
techniques with the factory default values one anatomical region at a time. Choose a
region and respond NO or YES.
DFLT CAL DAT allows the service personnel to reload the factory default calibration data
into the Power Module and Console. Because calibration data is involved, this screen
is protected by the XRAY CAL screen password. Enter the password and choose NO
or YES.
11.4 The POWER MODULE TO CONSOLE screen. (Memory card not
required)
The PM TO CONS screen will allow the installer to reload the site-specific calibration
data from the Power Module into a replacement Operator Console. This is required so
that the data stored in these two locations is in agreement. Select PM TO CONS,
enter the XRAY CAL password, and choose NO or YES.
LOAD DEFAULT
DFLT APR&CAL VET TECH
DFLT ALL APR
DFLT APR REG
DFLT CAL DAT