InnoVet Select
™
HF APR
Service Manual
16
1.5 Compatibility
Statement
This x-ray system is compatible with the following equipment:
X-RAY TUBES:
Toshiba E7239 (1.0/2.0 mm focal spots), Toshiba E7242 (0.6/1.5 mm focal spots),
Varian RAD 13 (1.0/2.0 mm focal spots), Varian RAD 14 (0.6/1.2 mm focal spots).
Contact the factory for tube compatibility not listed.
COLLIMATOR:
Summit G800 and D800, Progeny MC150 and Linear II or Linear IV
1.6 Accuracy
Statement
All measurements are made in compliance with the directions of the 20 kHz HF
APR Generator Service Manual in the Installation, Accuracy testing, and Calibration
procedures sections.
NOTE:
Specified accuracy for kVp and mA does not include test
equipment tolerance.
1.6.1 Exposure
kVp
Output kVp within ± 5% of indicated kVp, ± 1 kVp
1.6.2 Exposure
mAs
Output mAs within ± 10% of indicated mAs, ± 1 mAs
1.6.3 Duty
Cycle
The duty cycle is continuous for radiographic use. However, operation should be
limited to the capacity of the x-ray tube.