InnoVet Select
™
HF APR
Service Manual
46
Table 6-3: Table Interface Cable Connections at TB5 (continued)
NOTE:
The air switches are connected in parallel. The two wires (red) and (black) need
to connect to the TB5 terminal strip. See below
Origin Destination
Function
Air Switch cable - RED
TB5 - 1
Common
Air Switch cable - BLACK
TB5 - 4
Prep/Expose
Traditional Two-stage Footswitch - S1
TB5 - 2
Common
Traditional Two-stage Footswitch - S2
TB5 - 3
Prep
Traditional Two-stage Footswitch - S3
TB5 - 4
Expose
6.11 Interlocks
There are two possible interlocks besides the thermal interlock from the x-ray tube.
Terminals TB3-6 and TB3-8 are provided for connection of an isolated switch
contact for collimator interlock. If the collimation is manual, jumper these two
points.
Terminals TB3-5 and TB3-8 are provided for connection of an isolated switch
contact for door interlock. If no door interlock is being used, jumper these two
points.
TB3-7 to TB3-9 is a spare interlock.
TB3-8 and TB3-9 are electrically the same point and can be used
interchangeably.