InnoVet Select
™
HF APR
Service Manual
53
NOTE: Use the BACK button just below the console display to return to the
previous screen whenever necessary. From the CAL BOOTUP screen the
EXIT TO RAD command returns the system to normal operation.
8.3 XRAY CAL Screen
For all system configuration and generator calibration select XRAY CAL.
The console will require the installer to ENTER PASSWORD TO CONTINUE.
Press the outer most upper left, lower left, lower right and upper right arrow buttons
in a counter-clockwise sequence to enter the password.
WARNING:
Only personnel authorized to make system configuration selections and
perform generator calibration are to access the XRAY CAL screen.
Unauthorized changes by persons unfamiliar with the system will cause
equipment damage NOT COVERED BY WARRANTY!
The XRAY CAL screen will display as shown below.
TUBE SETUP:
To select x-ray tube type used at this installation while either keeping
existing calibration values or loading factory default calibration values
for this tube type.
PWR MOD SET:
To select kW, configure buckys/grid cabinets.
PWR MOD CAL:
To adjust independent mA leading edge points, master mA leading
edge for large and small spot, mA stabilized levels and master kVp.
AEC TBL SET:
Inactive
AEC WALL SET:
Inactive
FIL# TBL:
To view all filament numbers, which control mA leading edge.
MORE:
Factory settings only.
Press the BACK button to return to the CAL BOOTUP screen if desired.
XRAY CAL
TUBE SETUP
AEC WALL SET
PWR MOD SET
PWR MOD CAL
FIL# TBL
AEC TBL SET
MORE